RESEARCH

 

Field evaluation of a rapid, visually-read colloidal dye immunofiltration assay for Schistosoma japonicum for screening in areas of low transmission

 

Évaluation sur le terrain d'un test diagnostique rapide, à lecture visuelle, utilisant la technique d'immunofiltration et un colorant colloïdal, pour le dépistage de la schistosomiase sino-japonaise dans les zones de faible transmission

 

Evaluación sobre el terreno de una prueba visual rápida de inmunofiltración con colorante coloidal para Schistosomiasis japonica con fines de cribado en zonas de baja transmisión

 

 

Xiang XiaoI,II; Tianping WangII; Hongzhuan YeIII; Guangxiang QiangIV; Haiming WeiI; Zhigang TianV,1

IInstitute of Immunology, School of Life Sciences, University of Science and Technology of China, Hefei, Anhui, China
IIAnhui Institute of Schistosomiasis, Wuhu, Anhui, China
IIIWuhu Anti-Schistosomiasis Station, Anhui, China
IVNanling Anti-Schistosomiasis Station, Anhui, China
VInstitute of Immunology, School of Life Science, University of Science and Technology of China, Hefei, Anhui, 230027, China

 

 


ABSTRACT

OBJECTIVE: To determine the validity of a recently developed rapid test – a colloidal dye immunofiltration assay (CDIFA) – used by health workers in field settings to identify villagers infected with Schistosoma japonicum.
METHODS: Health workers in the field used CDIFA to test samples from 1553 villagers in two areas of low endemicity and an area where S. japonicum was not endemic in Anhui, China. All the samples were then tested in the laboratory by laboratory staff using a standard parasitological method (Kato–Katz), an indirect haemagglutination assay (IHA), and CDIFA. The results of CDIFA performed by health workers were compared with those obtained by Kato–Katz and IHA.
FINDINGS: Concordance between the results of CDIFA performed in field settings and in the laboratory was high (kappa index, 0.95; 95% confidence interval, 0.93–0.97). When Kato–Katz was used as the reference test, the overall sensitivity and specificity of CDIFA were 98.5% and 83.6%, respectively in the two villages in areas of low endemicity, while the specificity was 99.8% in the non-endemic village. Compared with IHA, the overall specificity and sensitivity of CDIFA were greater than 99% and 96%, respectively. With the combination of Kato–Katz and IHA as the reference standard, CDIFA had a sensitivity of 95.8% and a specificity of 99.5%, and an accuracy of 98.6% in the two areas of low endemicity.
CONCLUSION: CDIFA is a specific, sensitive, and reliable test that can be used for rapid screening for schistosomiasis by health workers in field settings.

Keywords: Schistosomiasis japonica/diagnosis; Immunoassay/methods; Colloids; Dyes; Hemagglutination tests; Parasite egg count/methods; Sensitivity and specificity; Comparative study; Evaluation studies; China (source: MeSH, NLM).


RÉSUMÉ

OBJECTIF: Déterminer la validité d'un test diagnostique rapide (CDIFA) récemment mis au point, faisant appel à un colorant colloïdal et à la technique d'immunofiltration et destiné au dépistage sur le terrain par des agents de santé de Schistosoma japonicum chez les villageois.
MÉTHODES: Des agents de santé présents sur le terrain ont appliqué le test CDIFA à des prélèvements effectués sur 1553 villageois provenant de deux zones de faible endémicité de la schistosomiase sino-japonaise et d'une zone où cette maladie n'est pas endémique, appartenant à la province d'Anhui, en Chine. Tous ces prélèvements ont ensuite été analysés en laboratoire par du personnel spécialisé utilisant une technique de diagnostic parasitologique classique (technique de Kato), la technique d'hémagglutination indirecte (IHA) et le test CDIFA. Les résultats des tests CDIFA effectués par les agents de santé ont été comparés à ceux obtenus par la technique de Kato et par hémagglutination indirecte.
RÉSULTATS: Le taux de concordance entre les résultats des tests CDIFA effectués sur le terrain et en laboratoire est élevé (indice kappa : 0,95, intervalle de confiance à 95 % : 0,93- 0,97). Si l'on utilise la technique Kato comme référence, on parvient à une sensibilité et à une spécificité globales pour le test CDIFA de 98,5 % et de 83,6 % respectivement dans les deux villages situés en zones de faible endémicité, ainsi qu'à une spécificité de 99,8 % dans le village exempt de schistosomiase sino-japonaise à l'état endémique. Par comparaison avec la technique d'hémagglutination indirecte (IHA), la spécificité et la sensibilité globales du test CDIFA sont respectivement supérieures à 99 et 96 %. Si l'on se réfère à une combinaison de la technique KATO et de la technique d'hémagglutination indirecte, on obtient pour le test CDIFA une sensibilité de 95,8 % et une spécificité de 99,5 %, ainsi qu'une précision de 98,6 % dans les deux zones d'endémicité.
CONCLUSION: Le test CDIFA offre une méthode spécifique, sensible et fiable permettant le dépistage rapide sur le terrain de la schistosomiase par les agents de santé.

Mots clés: Schistosomiase artérioveineuse/diagnostic; Dosage immunologique/méthodes; Colloïde; Colorants; Réaction hémagglutination; Numération œuf parasite/méthodes; Sensibilité et spécificité (Epidémiologie); Etude comparative; Etude évaluation; Chine (source: MeSH, INSERM).


RESUMEN

OBJETIVO: Determinar la validez de una prueba rápida desarrollada recientemente -prueba de inmunofiltración con colorante coloidal (PICC)- y empleada por agentes de salud sobre el terreno para identificar a los lugareños infectados por Schistosoma japonicum.
MÉTODOS: Agentes de salud sobre el terreno utilizaron la PICC para analizar muestras de 1553 lugareños en dos zonas de baja endemicidad y una zona sin S. japonicum endémica de Anhui, China. Todas las muestras fueron analizadas en el laboratorio por personal técnico utilizando un método parasitológico estándar (Kato-Katz), una prueba de hemaglutinación indirecta (IHA) y la PICC. Los resultados de la PICC llevada a cabo por los agentes de salud fueron comparados con los obtenidos mediante las técnicas de Kato-Katz e IHA.
RESULTADOS: Se observó una elevada concordancia entre los resultados de la PICC aplicada sobre el terreno y la practicada en el laboratorio (índice kappa = 0,95; intervalo de confianza del 95% = 0,93–0,97). Cuando se utilizó el método Kato-Katz como prueba de referencia, la sensibilidad y la especificidad generales de la PICC fueron de un 98,5% y un 83,6%, respectivamente, en las dos aldeas de las zonas de endemicidad baja, mientras que la especificidad fue del 99,8% en la aldea sin endemicidad. En comparación con la IHA, la especificidad y la sensibilidad generales de la PICC superaron el 99% y el 96%, respectivamente. Utilizando la combinación de Kato-Katz e IHA como referencia, la PICC presentó una sensibilidad del 95,8% y una especificidad del 99,5%, con una precisión del 98,6%, en las dos zonas de endemicidad baja.
CONCLUSIÓN: La PICC constituye una prueba específica, sensible y fiable, que se presta a ser utilizada por los agentes de salud sobre el terreno para el cribado rápido de la esquistosomiasis.

Palabras clave: Esquistosomiasis japonica/diagnóstico; Inmunoensayo/métodos; Coloides; Tinturas; Tests de hemaglutinación; Recuento de huevos de parásitos/métodos; Sensibilidad y especificidad: Estudio comparativo; Estudios de evaluación; China (fuente: DeCS, BIREME).



 

 

Introduction

Schistosomiasis remains a public health problem in China, although great advances in schistosomiasis control have been achieved during the past five decades (1). It has been estimated that in 2002 more than 0.8 million people were infected, and about 30 million people were at risk of infection (1, 2). At present, selective chemotherapy is the main strategy for control of schistosomiasis in China because the prevalence and the morbidity of the disease are low after many years of mass chemotherapy (1, 3). Thus the "test-treat" approach is probably the most cost-effective strategy for identifying individuals to be targeted for repeated treatments (4). For this approach to be adopted by the schistosomiasis control programme in China, there is an essential need for a simple, rapid, sensitive and inexpensive screening technique.

Some indirect methods, including interpretation of clinical symptoms, ultrasound examination, measurement of markers of biochemical or morbidity, and the application of questionnaires, have been proved useful in screening for Schistosoma haematobium in Africa and are cost-effective in areas where the infection is highly endemic (5, 6). However, these methods are of very limited usefulness in screening for intestinal schistosomiasis caused by S. mansoni or S. japonicum, especially in areas of low or declining prevalence, because of their lack of specificity (7, 8).

Direct parasitological techniques, apart from being labour-intensive and time-consuming, have become relatively insensitive owing to the widespread use of chemotherapy resulting in generally lower worm burdens per individual. A definitive diagnosis of infection with S. japonicum still relies on the identification of eggs or miracidia in the faeces. In order to improve the identification of individuals to be targeted for treatment, however, immunodiagnostic methods such as the circumoval precipitin test (COPT), indirect fluorescence assay (IFA) and enzyme-linked immunosorbent assay (ELISA), which have a high sensitivity, specificity, and reliability, and high compliance compared with parasitological techniques, have been integrated into the control programme in China since the early 1980s when the safe and effective anti-schistosomiasis drug (praziquantel) was introduced (4, 9). These immunodiagnostic tests share drawbacks, however, including cost, requirements for laboratory equipment and specific expertise, electricity, refrigeration, and incubation periods that cause a delay in diagnosis, which make them unsuitable for direct use in field settings. The lack of effective tools for screening has been recognized as a major barrier to the prevention and control of this infection in areas of low transmission.

We have recently developed a novel method for rapid screening for infection with S. japonicum – a colloidal dye immunofiltration assay (CDIFA) that is based on the detection of antibody to the heat-resistant, soluble egg antigen of S. japonicum. Laboratory studies have shown that CDIFA is a rapid, convenient and reliable method for screening for schistosomiasis infection (10). In the study reported here, we further evaluated the performance and acceptability of CDIFA in field situations and compared it with the indirect haemagglutination test (IHA) and Kato–Katz method.

 

Materials and methods

Study site and participants

The study was conducted between August and November 2003 in three widely separated areas of Anhui, China: Dongliang village (village A) in Wuhu City, Puxi village (village B), in Lanling County, and Dayao village (village C, a non-endemic area) in Shouxian County. All the subjects were asymptomatic villagers who were self-selected to participate in a mass screening survey on parasitic diseases. All participants in the three villages were included in this study, except those people in village C who had ever visited an endemic area (Fig.1). Age, sex, schistosomiasis history and data on treatment were recorded for all participants.

 

 

Specimen collection and handling

In all three studies, field health workers were blinded to the schistosomiasis status of the subject to be tested. Samples of 250 µl of ear lobe blood were collected using the heparinized plastic capillary tubes (2 mm in diameter) provided with the CDIFA kit, sealed at one end. The tubes were then maintained in a vertical position for about 1 h to allow the spontaneous separation of the plasma and red blood cells. Plasma was removed by cutting the tubes with scissors. The plasma was tested immediately with the CDIFA kit on the spot, and the residual plasma, as well as a faecal sample obtained from the same subject, was taken to local schistosomiasis stations where CDIFA, IHA and Kato–Katz tests were performed by laboratory staff within 24 h of sample collection.

Stool examinations

The Kato–Katz thick smear stool examination (11) was performed to identify and quantify S. japonicum eggs in the faeces of each subject. The stool examination was based on three slides (41.7 mg per smear) prepared from a single stool specimen, resulting in a total sample weight of 125 mg. Slides were read 12–48 h after their initial preparation by two experienced technicians who were unaware of the subject's medical status or immunological test result. S. japonicum egg counts were expressed in eggs per gram (EPG).

Repeat stool examination

In village B, 40 randomized individuals all with a negative result by initial Kato–Katz test and half of them with a positive result by CDIFA were selected for repeat stool examination. Three repeated faecal samples from each person on three separate days were collected and tested by Kato–Katz technique (10 slides from a single stool sample) and egg hatching test (12, 13).

Indirect haemagglutination test

IHA was performed as described in previous reports (12, 14). The titre in the test plasma was recorded as one dilution before that which yielded a clear, sharp dark spot similar to that in the negative control wells. Titres were expressed as reciprocal values. Titres of > 10 were considered to indicate a positive result (12, 14).

CDIFA

CDIFA was performed as described previously (10). Standard operating procedures are as follows: (i) add two drops of washing buffer from the buffer bottle to a well on the test card; (ii) add one drop of plasma taken directly from the blood collecting tube to the same well, and allow it to be absorbed completely; (iii) add two drops of dye-labelled soluble egg antigen conjugate from the detecting bottle to the well, and allow to soak in; (iv) add two drops of washing buffer to the well to remove the unbound conjugate, and then read the result immediately. The appearance of two red dots in the well indicates a positive reaction, and the appearance of a single red dot indicates a negative reaction. The assay is performed at room temperature without incubation and the whole procedure can be completed within 2 min.

Training

The CDIFA kits were stored at room temperature (12–30 ºC) during the study. Five laboratory technicians, five medical doctors and 12 health workers participated in running and reading CDIFAs over the 12 weeks of the study. All received 2 h of training on site from one of the authors (X.X. or W.T.) in how to obtain samples and to use, read, and archive the assay.

CDIFA was performed on fresh plasma in the field. Two aliquots (one for CDIFA and one for IHA) were made from the residual plasma and each aliquot was assigned a randomly-generated code to ensure that the technicians taking part in the study in laboratory could not trace the identity of the subject giving the sample.

Questionnaire

To assess ease of use of the three assays, the technicians and health workers involved were asked to complete a questionnaire on their impressions of test presentation, simplicity of the test procedure, and ease of interpretation of test results.

Data analysis

Only data from subjects who accepted all three tests was used for the analyses, which were performed using the SPSS Statistical Package for Social Sciences, Version 11.5 software (SPSS Inc., Chicago, IL, 2002). To measure test agreement between field health workers and laboratory staff, the kappa index was calculated. Owing to the fact that the test was expected to be performed mainly by field health workers rather than by laboratory staff in the future, only the results by field health workers were use to calculate sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV), with microscopy and/or IHA results being used as the "reference standards". The variables measured were the number of true positives (TP), number of true negatives (TN), number of false positives (FP), and number of false negatives (FN). Sensitivity was calculated as TP/(TP + FN), specificity was calculated as TN/(TN + FP), the PPV was calculated as TP/(TP + FP), and the NPV was calculated as TN/(FN + TN). Test accuracy – the proportion of all tests that gave a correct result – was defined as (TP + TN)/(number of all tests). Reliability was expressed as the Youden index (sensitivity + specificity –1).

 

Results

Test results

Of the 1598 participants, 45 were excluded from the analysis owing to incomplete data (they were not able to offer stool samples) (Fig.1). The vast majority of the subjects excluded were from village C, with only four being from village A and one from village B. All the excluded subjects had negative results for both CDIFA and IHA. The initial test results and the characteristics of the remaining 1553 subjects are shown in Table 1. The geometric mean egg count for infected persons was 8 eggs per gram (EPG) in village A, and 15.6 EPG (range, 8–280 EPG) in village B.

 

 

Concordance between results obtained by field health workers and laboratory staff

The agreement between the results of CDIFA tests performed by field health workers or by laboratory staff was excellent. Only 21 (1.4%) of the results of 1553 tests by both groups were discordant; of these, 12 positive results obtained by field health workers were declared to be negative by laboratory staff, and the remaining 9 negative results were declared to be positive by laboratory staff. For all 1553 samples, the kappa index was 0.95 (95% confidence interval (CI), 0.93–0.97).

Performance of CDIFA and IHA compared with microscopy

The results of CDIFA and IHA compared with those of microscopy are shown in Table 2 and Table 3.

 

 

 

 

In the population of village A, CDIFA had a sensitivity of 100% and a specificity of 92% compared with microscopy. The PPV of this test was 4.8% and the NPV was 100%. In comparison, IHA had a sensitivity of 100% and a specificity of 91.8%, with a PPV of 4.7% and NPV of 100%.

In the population of village B, CDIFA performed with a sensitivity of 98.5% and a specificity of 69.6% compared with microscopy, with PPV 27.7% and NPV 99.7%. In comparison, IHA showed a sensitivity of 98.5% and a specificity of 68.7%, while the PPV was 27.1% and the NPV was 99.7%.

In the population of village C, a non-endemic area, both CDIFA and IHA showed excellent specificity (99.6% and 98.79%, respectively). The agreement between CDIFA and microscopy was 92.0% in village A, 72.6% in village B and 99.8% in village C.

Performance of CDIFA compared with IHA

The results of screening for S. japonicum using CDIFA compared with IHA (the most commonly used immunological method for schistosomiasis screening in China) are shown in Table 4. In the two villages in which S. japonicum was endemic, CDIFA had a sensitivity of 96.1% and a specificity of 99.4% compared with IHA, while the PPV was 98.2% and the NPV was 98.7%. The agreement between CDIFA and IHA was 99% in village A, 98.2% in village B and 99.5% in village C.

 

 

Performance of CDIFA compared with the combination of microscopy and IHA

If the combination of microscopy and IHA was taken as the "reference standard" (which was not the case), the overall sensitivity and specificity of CDIFA were 95.8% (95% CI, 93.4–98.1%) and 99.5% (95% CI, 99.1–100%) in the two villages of low endemicity, respectively, while the PPV was 98.2% (95% CI, 96.6–99.8%) and the NPV was 98.6% (95% CI, 97.1–100%), with an accuracy of 98.6% (95% CI, 97.9–99.4%).

Performance of CDIFA compared with repeat stool examination

Of the 20 subjects who had positive results by CDIFA, 10 were finally confirmed as positive for schistosomiasis by three repeat stool examinations; 6 of these subjects had no history of schistosomiasis and 4 had reported recent treatment (8 months previously) with praziquantel. Of the other 20 subjects who had negative results by CDIFA, none was finally confirmed as having a positive result.

Assay characteristics and ease of use

Minimal equipment, such as an incubator, microplate shaker and precision pipettes are needed to perform IHA. CDIFA, however, requires no specialist instruments or electricity, because the kit contains all the necessary equipment and reagents, and the reagents are stable at room temperature for more than 6 months (10). The testers' impressions of the presentation of the test, simplicity of the operating procedure, and ease of interpretation of results for CDIFA, IHA and the Kato–Katz test are shown in Fig. 2. CDIFA was reported to be the best presented, and easiest to use and to interpret of these three tests.

 

 

Discussion

The term "ASSURED" ("affordable, sensitive, specific, user-friendly, rapid and robust, equipment-free and deliverable to end-users") has been coined to describe the ideal characteristics of a diagnostic test for use in developing countries (15). For a screening test to be used in areas where schistosomiasis is of low endemicity, sensitivity is of course given top priority (4, 9). The present study showed that the sensitivity of CDIFA is excellent, being greater than 98% in the two villages of low endemicity (villages A and B) studied. However, the specificity of CDIFA in village B (about 70% compared with the Kato–Katz test) may not be adequate, since large numbers of uninfected persons (but who had a positive test result by CDIFA) would be given chemotherapy. The low specificity of CDIFA was not caused by cross-reaction with intestinal nematodes, which is often the case with other routine serological tests such as enzyme-linked immunoabsorbent assay (ELISA) tests that use crude antigen (9); for example, in the non-endemic village, the specificity of CDIFA was excellent despite the fact that 24% of residents were infected with Ascaris, 12% with hookworm and 9% with Trichuris (data not shown). In addition, the prevalence of intestinal nematodes was similar in the three villages (data not shown) and no cases of infection with Paragonimus or Clonorchis were found in this study. This is in agreement with the results of studies with CDIFA in the laboratory that show no cross-reaction with intestinal nematodes (10).

In our experience with sero-epidemiological studies in different regions of China, we have found that rates of seroprevalence of S. japonicum are 2–4 times higher than rates of prevalence determined by methods relying on faecal parasitology. This difference seems to be caused mainly by the poor sensitivity of the parasitological methods and thus failure to detect infection in some patients with low egg counts (4, 9, 16, 17). The high sensitivity of CDIFA and the capacity to detect very low-level infections that would not be detected by faecal examination might explain the difference observed between the serological and parasitological methods investigated in this study. Yu et al. (18) reported that more than half of the infected people in endemic areas (i.e. areas with a S. japonicum prevalence of at least 26%) would fail to be detected by duplicate Kato–Katz smears (41.7 mg of faecal sample each). Many epidemiological studies carried out in areas of low endemicity in China (19), or other countries (20–22), have given similar results. In this study, half of the 20 randomized subjects with a positive result by CDIFA but a negative result by Kato–Katz test in the first survey were finally confirmed as infected by three repeated stool examinations. It is clear that the true specificity and PPV of CDIFA for the detection of schistosomiasis is higher than that obtained when using a single Kato–Katz stool examination as the gold standard.

The seroprevalence of S. japonicum may also be higher than the prevalence determined by methods relying on faecal parasitology because of the inability of the serological tests to discriminate between present and past infections (4, 9). Many studies suggest that, in areas of low endemicity, results obtained by serology more closely approximate the true situation than do results obtained by parasitology (9, 19, 23, 24). For schistosomiasis control programmes, it is important to use the method that most accurately estimates the true prevalence of schistosomiasis. This is important not only for a better epidemiological understanding, but also for more efficient coverage with chemotherapy in control programmes. Specific antibodies to schistosomiasis may disappear within 1 to 3 years, or after a period 5 or 7 years, depending on the technique and antigen used (9, 25, 26). However, the antigen used in CDIFA – purified heat-resistant soluble egg antigen – differs from crude soluble egg antigen and the antigens used in conventional immunological methods, and the question of at what time-point the results of CDIFA become negative in successfully treated patients needs to be addressed.

For instrument-independent tests, reproducibility is another intrinsic factor that may affect performance. Operator variability for CDIFA was 1.4% between qualified laboratory staff and minimally trained field health workers, illustrating its excellent performance. The results of the questionnaire demonstrated the preference of the operators for CDIFA (Fig. 2).

In addition to sensitivity, specificity, simplicity and stability, some other factors, such as kit costs, should also be considered in the choice of a screening assay for application in field settings. The cost per test for CDIFA is very low, about US$ 0.12; however, many other factors can also affect cost-effectiveness. Therefore CDIFA should be incorporated into community control programmes to analyse its cost-effectiveness in areas where S. japonicum is endemic at different levels of prevalence.

Whether a screening test is suitable for a specific task depends on its predictive value in given settings, in particular, according to the prevalence of the condition to be detected. On the basis of the sensitivity and specificity as determined by field health workers, we showed that, with the testing strategies applied, CDIFA had a PPV and NPV both of greater than 98% in areas of low endemicity, compared with the combination of Kato–Katz and IHA. This is particularly important in selective chemotherapy strategies in areas of low endemicity where a negative test result needs to predict a true negative case with a probability of close to 100%. Considering the fact that the work described here was carried out independently in remote villages by field health workers with minimal training, the performance of CDIFA was excellent, suggesting that CDIFA can usefully be employed as a sensitive and reliable screening test for S. japonicum.

 

Acknowledgements

We wish to thank Professor Wu Weidui and other colleagues from the Anhui Institute of Parasitic Diseases and the Wuhu Anti-Schistosomiasis Station, who helped greatly with the field research work. We also express our appreciation for the skill and dedication of the field health workers deployed by the Nanling Anti-Schistosomiasis Station, and the Wuhu Anti-Schistosomiasis Station. We also would like to thank Professor M.G. Taylor, London School of Hygiene and Tropical Medicine, for his valuable comments on the manuscript.

Funding: This work was supported partly by an Outstanding Young Scientist Award (No. 30125038) and Key Project (No. 30230340) of the Natural Science Foundation of China to ZT, and Key Science Projects from Anhui Province (No. 01303009) to XX.

Competing interests: none declared.

 

References

1. Jiang QW, Wang L, Guo JG, Chen M, Zhou XN, Engels D. Morbidity control of schistosomiasis in China. Acta Tropica 2002;82:115-25.        

2. Chen XY, Wu X, Wang LY, Dang H, Wang Q, Zhen J, et al. Schistosomiasis situation in the People's Republic of China in 2002. Chinese Journal of Schistosomiasis Control 2003;15,241-5.        

3. Zhang SJ, Ling DD. The potential risk and control strategy in low endemic area of schistosomiasis in China. Acta Tropica 2002;82:289-93.        

4. Wu GL. A historical perspective on the immunodiagnosis of schistosomiasis in China. Acta Tropica 2002;82:193-8.        

5. Lengeler C, Utzinger J, Tanner M. Questionnaires for rapid screening of schistosomiasis in sub-Saharan Africa. Bulletin of the World Health Organization 2002;80:235-42.        

6. Mafe MA. The diagnostic potential of three indirect tests for urinary schistosomiasis in Nigeria. Acta Tropica 1997;68:277-84.        

7. Barreto ML. Questionnaire approach to diagnosis in developing countries. Lancet 1998;352:1164-5.        

8. Lengeler C, Utzinger L, Tanner M. Screening for schistosomiasis with questionnaires. Trends in Parasitology 2002;18:375-7.        

9. Li Y. Advance of the study on immunodiagnosis of schistosomiasis. In: Li Y, editor. Immunology and immunodiagnosis of parasitic diseases. Nanjing, China: Jiangsu Science Technology Publishing House; 1991. p. 210-34.        

10. Xiao X, Wang TP, Tian ZG. Development of a rapid, sensitive, dye immunoassay for schistosomiasis diagnosis: a colloidal dye immunofiltration assay. Journal of Immunological Methods 2003;280:49-57.        

11. Katz N, Chaves A, Pellegrino J. A simple device for quantitative stool thick-smear technique in schistosomiasis mansoni. Revista do Instituto de Medicina Tropical de Sao Paulo 1972;14:397-400.        

12. Department of Diseases Control, Ministry of Health. Textbook for schistosomiasis control. 2nd ed. Shanghai: Shanghai Publishing House for Sciences and Technology; 2000. pp. 84-91.        

13. Hubbard A, Liang S, Maszle D, Qiu D, Gu X, Spear RC. Estimating the distribution of worm burden and egg excretion of Schistosoma japonicum by risk group in Sichuan Province, China. Parasitology 2002;125:221-31.        

14. Song Y, Xiao S, Wu W, Zhang S, Xie H, Xu X, et al. Preventive effect of artemether on schistosome infection. Chinese Medical Journal 1998;111:123-7.        

15. Mabey D, Peelin RW, Ustianowsk A, Perkin MD. Tropical infectious diseases: diagnostics for the developing world. Nature Reviews Microbiology 2004;2:231-40.        

16. Engels D, Sinzinkayo E, Gryseels B. Intraspecimen fecal egg count variation in Schistosoma mansoni infection. The American Journal of Tropical Medicine and Hygiene 1997;57:571-7.        

17. Engels D, Sinzinkayo E, Gryseels B. Day-to-day egg count fluctuation in Schistosoma mansoni and its operational implications. The American Journal of Tropical Medicine and Hygiene 1996;54:319-24.        

18. Yu JM, De Vlas SJ, Yuan HC, Gryseels B. Variations in fecal Schistosoma japonicum egg counts. The American Journal of Tropical Medicine and Hygiene 1998;59:370-5.        

19. Yu JM, Yuan HC, Chen Q, Yang Q, Zhang SJ, Jiang QW. Comparative study on detection of schistosomiasis infection among repeated Kato–Katz method, IHA and stool hatching test. Chinese Journal of Schistosomiasis Control 1997;9:150-3.        

20. De Vlas SJ, Gryseels B. Underestimation of Schistosoma mansoni prevalences. Parasitogy Today 1992;8:274-7.        

21. Kongs A, Marks G, Verle P, Van der Stuyft P. The unreliability of the Kato–Katz technique limits its usefulness for evaluating S. mansoni infections. Tropical Medicine and International Health 2001;6:163-9.        

22. Utzinger J, Booth M, N'Goran EK, Muller I, Tanner M, Lengeler M, et al. Relative contribution of day-to-day and intra-specimen variation in faecal egg counts of Schistosoma mansoni before and after treatment with praziquantel. Parasitology 2001;122,537-44.        

23. Dias LCS, Kanamura HY, Hoshino-Shimizu S. Field trials for immunodiagnosis with reference to Schistosoma mansoni. In: Bergquist, NR, editor. Immunodiagnostic approches in schistosomiasis. Chichester: John Wiley; 1992. p. 39-47.        

24. Kanamura,HY, Dias LCS, da Silva RM, Glasser CM, Patucci RM, Vellosa SA, et al. A comparative epidemiologic study on specific antibodies (IgM and IgA) and parasitological findings in an endemic area of low transmission of S. mansoni. Revista do Instituto de Medicina Tropical de Sao Paulo 1998;40:85-91.        

25. Zhu YC, Hua W, Liu Y, He W, Xu Y, Jiang Y. A study on evaluation of efficacy of chemotherapy for schistosomiasis with fraction antigen of soluble egg antigen of Schistosoma japonicum. Chinese Journal of Schistosomiasis Control 1996;8:321-4.        

26. Doenhoff MJ, Chiodini PL, Hamilton JV. Specific and sensitive diagnosis of schistosome infection: can it be done with antibodies? Trends in Parasitology 2004;20:35-9.        

 

 

(Submitted: 26 August 2004 – Final revised version received: 26 January 2005 – Accepted: 26 January 2005)

 

 

1 Correspondence should be sent to this author (email: ustctzg@yahoo.com.cn).

World Health Organization Genebra - Genebra - Switzerland
E-mail: bulletin@who.int